Objective:To establish a method for determination of methyl tert-butyl ether(MTBE) in water by static state headspace gas chromatography.Methods:Static state headspace gas chromatography was used to determine MTBE in water.MTBE was separated by ploar capillary column.Results:The detection limit of MTBE in water was 0.0163 μg/L,the linear range 0.0156~120 mg/L.The coefficient of correlation was 0.9997,recoveries 92.6%~100.1% and RSD less than 1.1%.Conclusion:...