On the basis of accurate HPGe detector's size which is obtained by CT scanning, the calculation of point source physical model in HPGe gamma spectrometer is established by using MCNP4C Monte Carlo code; By means of correcting the detector dead layer, the final dead layer thickness is identified as 0.38 cm; Based on the point source physical model and the result of detector dead layer correction, detection efficiencies of each energy branch of 60Co, 133Ba, 137Cs, 241Am and 152Eu standard point sources are calculated by MCNP4C, comparing with the...